Dako Hybridizer Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Outils Dako Hybridizer. Dako Hybridizer User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Handbook

Dako Hybridizer | Handbook 1 Hybridizer | Handbook

Page 2 - Hybridizer

10 Dako Hybridizer | Handbook Section 2 | System Overview Principle and Intended Use For in vitro diagnostic use to co-denature and hybridize

Page 3

Dako Hybridizer | Handbook 11 Symbols and Definitions (cont.) Display Abbreviations PGM Program Denat & Hyb Denaturation & Hybridization

Page 4 - Table of Contents

12 Dako Hybridizer | Handbook Audible Indicators Single beep – on all legal keystrokes. Two quick beeps – upon accepting a field and cursor has mov

Page 5

Dako Hybridizer | Handbook 13 High ambient temperature condition: The instrument will attempt to achieve process set temperatures. However, if

Page 6 - How to use this handbook

14 Dako Hybridizer | Handbook If the ambient temperature changes during a process and causes the instrument process set temperature to change beyon

Page 7

Dako Hybridizer | Handbook 15 If the ambient temperature changes after a process is completed, but before the user removes the slides and causes

Page 8 - Connect Power

16 Dako Hybridizer | Handbook Section 3 | Programming Opening and Closing the Lid The plate may be hot. Use caution and check temperature on di

Page 9

Dako Hybridizer | Handbook 17 Note: If all 40 program numbers have been used, the “Create a PGM” line in the Main Menu will no longer appear. An

Page 10 - Section 2

18 Dako Hybridizer | Handbook Create a Hybridization Only Program (Hyb Only) Turn unit on and wait for the Main Menu screen. Use the arrows keys to

Page 11

Dako Hybridizer | Handbook 19 Create a Fixed Temperature Program (Fixed Temp) Use the arrows keys to move cursor to “Create a PGM” line and pres

Page 12 - Error Messages

2 Dako Hybridizer | Handbook Hybridizer | Handbook Document Number 0003061 Revision C July 2007 For In Vitro Diagnostic Use

Page 13

20 Dako Hybridizer | Handbook Section 4 | Operating Instructions Opening and Closing the Lid The plate may be hot. Use caution and check temp

Page 14

Dako Hybridizer | Handbook 21 Display will confirm PGM number/name and Denat & Hyb times and temperatures. Cursor highlights “Run PGM” line.

Page 15

22 Dako Hybridizer | Handbook The Hybridizer will automatically cool to hybridization set temperature once denaturation is completed. “Ambient

Page 16 - Section 3

Dako Hybridizer | Handbook 23 Display will confirm PGM number/name and Hyb Only times and temperatures. Cursor highlights the “Run PGM” line.

Page 17

24 Dako Hybridizer | Handbook Upon program completion Hybridizer will beep five times and the display will show “Process Complete”. Hybridization t

Page 18

Dako Hybridizer | Handbook 25 When the fixed temp set temperature is reached the display prompts to “Add Slides and Close Lid”. Before adding sli

Page 19

26 Dako Hybridizer | Handbook Recommended for Best Results  Place glass coverslip over probe and sample  Apply sealant along all coverslip edge

Page 20 - Opening and Closing the Lid

Dako Hybridizer | Handbook 27 Predefined Limits Program Mode Temperature Range Timer Limits Denature 50°C to 99°C 0-30 minutes Hybridization Ro

Page 21 - - Close Lid Start

28 Dako Hybridizer | Handbook Section 5 | Edit a Program Opening and Closing the Lid The plate may be hot. Use caution and check temperature

Page 22 - Run a Hyb Only Program

Dako Hybridizer | Handbook 29 Cursor highlights existing program type: “Denat & Hyb”, “Hyb only” or “Fixed Temp”. Press “Enter” button to acc

Page 23 - - Close Lid

Dako Hybridizer | Handbook 3 Copyright© 2007 Dako Colorado, Inc. All rights reserved This document may not be copied in whole or in part or re

Page 24 - Run a Fixed Temp Program

30 Dako Hybridizer | Handbook Section 6 | Maintenance Overview Dako recommends that instrument operators perform periodic inspections and preve

Page 25 - Abort Program in Process

Dako Hybridizer | Handbook 31 Troubleshooting Unit does not turn on, no power. Check both cord ends are plugged in. Check Fuses located on rear

Page 26 - Humidity Control Strips

32 Dako Hybridizer | Handbook Appendix A | Specifications Code Numbers S 2450 S 2451 Electrical 100-120 VAC @ 3.0 A 220-240 VAC @ 1.6A P

Page 27 - Predefined Limits

4 Dako Hybridizer | Handbook Table of Contents | Handbook Section 1 | Unpacking and Installation...

Page 28 - Section 5

Dako Hybridizer | Handbook 5 Section 5 | Edit a Program... 28 O

Page 29

6 Dako Hybridizer | Handbook How to use this handbook This handbook along with information contained on product labels should provide all the inf

Page 30 - Section 6

Dako Hybridizer | Handbook 7 This page is intentionally blank.

Page 31 - Troubleshooting

8 Dako Hybridizer | Handbook Section 1 | Unpacking and Installation Inspect Packaging The Hybridizer and its accessories are delivered in one c

Page 32 - Appendix A

Dako Hybridizer | Handbook 9 This page is intentionally blank.

Commentaires sur ces manuels

Pas de commentaire